User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Philips PM 6685 DATE: 19-Feb-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 65 NUMBER OF LINES: 301 CONFIGURATION: Fluke 6060B ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F W 1.002 HEAD INTRODUCTION 1.003 DISP This procedure is intended to: 1.003 DISP - Check the instrument's specification. 1.003 DISP - Check the necessity of re-calibration. 1.004 OPBR Does the UUT have a IEEE option (PM 9626)? 1.005 JMPT 33.002 1.006 DISP UUT Setup: Press PRESET, then ENTER 1.006 DISP 1.006 DISP Connect UUT Ext Ref input (Rear) to House standard. 1.006 DISP Connect UUT Ch A input to 6060 RF output. 1.006 DISP 1.007 6060 10MH 0.707V -D S 1.008 OPBR Does the UUT have a PM 9678B (02 option) ? 1.009 JMPF 2.002 1.010 MEMI Please enter reading (Example: XX.XXXXXXXX) 1.011 MATH MEM=MEM*1E6 1.012 MEMC 10E6H 15.0U 2.001 JMP 5.001 2.002 OPBR Does the UUT have a PM 9690 (04 option) ? 2.003 JMPF 3.002 2.004 MEMI Please enter reading (Example: XX.XXXXXXXX) 2.005 MATH MEM=MEM*1E6 2.006 MEMC 10E6H 2.00U 3.001 JMP 5.001 3.002 OPBR Does the UUT have a PM 9691 (05 option) ? 3.003 JMPF 4.002 3.004 MEMI Please enter reading (Example: XX.XXXXXXXX) 3.005 MATH MEM=MEM*1E6 3.006 MEMC 10E6H 1.00U 4.001 JMP 5.001 4.002 MEMI Please enter reading (Example: XX.XXXXXXXX) 4.003 MATH MEM=MEM*1E6 4.004 MEMC 10000000H 150U 5.001 HEAD {Sensitivity and Frequency Range Tests} 5.002 DISP UUT Setup: 5.002 DISP Select IMP 50 ohms, Maximum Sensitivity, Ext Ref on. 5.003 MESS 5.004 MESS Edit 6060 Level Up/Down to Find the stable trigger 5.004 MESS point. 5.005 6060 10mV +50% 1MH -D 6.001 6060 10mV +50% 25MH -D 7.001 6060 10mV +50% 50MH -D 8.001 6060 15mV +50% 100MH -D 9.001 6060 20mV +50% 150MH -D 10.001 6060 30mV +50% 200MH -D 11.001 6060 50mV +50% 300MH -D 12.001 HEAD {Measuring Functions} 12.002 6060 10MH 0.707V -D S 12.003 MESS UUT Setup: Select PRESET, ENTER 12.003 MESS 12.003 MESS Connect 6060 to Ch A & Ch E (Ext Arm on rear) 12.003 MESS using a BNC Tee. 12.003 MESS 12.004 EVAL Does {UUT Display 10 Mhz (Preset)} ? 13.001 MESS UUT Setup: Select IMP A 50 ohms 13.001 MESS 13.002 6060 10MH 0.707V -D S 13.003 EVAL Does {UUT Display 10 Mhz (IMP A 50)} ? 14.001 MESS UUT Setup: Select Non AUTO 14.001 MESS 14.002 6060 10MH 0.707V -D S 14.003 EVAL Does {UUT Display 10 Mhz (Non AUTO)} ? 15.001 MESS UUT Setup: Select PER A 15.001 MESS 15.002 6060 10MH 0.707V -D S 15.003 EVAL Does {UUT Display 100 nS (PER A)} ? 16.001 MESS UUT Setup: Select RATIO A/E 16.001 MESS 16.002 6060 10MH 0.707V -D S 16.003 EVAL Does {UUT Display 1.0000000 (RATIO A/E)} ? 17.001 MESS UUT Setup: Select PWIDTH A 17.001 MESS 17.002 6060 10MH 0.707V -D S 17.003 EVAL Does {UUT Display 50 nS (PWIDTH A)} ? 18.001 MESS UUT Setup: Select TOT A MAN, DISPLAY HOLD 18.001 MESS 18.002 6060 10MH 0.707V -D S 18.003 EVAL Does {UUT Start counting (TOT A MAN, DISPLAY HOLD)} ? 19.001 MESS UUT Setup: Select DISPLAY HOLD 19.001 MESS 19.002 6060 10MH 0.707V -D S 19.003 EVAL Does {UUT Stop counting (DISPLAY HOLD)} ? 20.001 MESS UUT Setup: Select DUTY FACTOR 20.001 MESS 20.002 6060 10MH 0.707V -D S 20.003 EVAL Does {UUT Display 0.500000 (DUTY FACTOR)} ? 21.001 MESS UUT Setup: Select AUTO 21.001 MESS 21.002 6060 10MH 0.707V -D S 21.003 EVAL Does {UUT Display 0.500000 (AUTO)} ? 22.001 HEAD Channel C Options 22.002 OPBR Does the UUT have a Channel C option? 22.003 JMPF 32.001 22.004 OPBR Does the UUT have a 1.3 Ghz channel C option (PM 9621)? 22.005 JMPT 22.010 22.006 OPBR Does the UUT have a 2.7 Ghz channel C option (PM 9624)? 22.007 JMPT 26.002 22.008 OPBR Does the UUT have a 4.5 Ghz channel C option (PM 9625)? 22.009 JMPT 29.002 22.010 HEAD {PM 9621 Channel C option test} 22.011 DISP UUT Setup: 22.011 DISP Connect UUT Channel C input to 6060 RF output. 22.012 MESS 22.013 MESS Edit 6060 Level Up/Down to Find the stable trigger 22.013 MESS point. 22.014 6060 10mV +50% 70MH -D 23.001 6060 15mV +50% 900MH -D 24.001 6060 15mV +50% 1100MH -D 25.001 6060 40mV +50% 1300MH -D 26.001 JMP 32.001 26.002 HEAD {PM 9624 Channel C option test} 26.003 DISP UUT Setup: 26.003 DISP Connect UUT Channel C input to 6060 RF output. 26.004 MESS 26.005 MESS Edit 6060 Level Up/Down to Find the stable trigger 26.005 MESS point. 26.006 6060 20mV +20U 100MH -D 27.001 6060 10mV +50% 300MH -D 28.001 6060 10mV +50% 2100MH -D 29.001 JMP 32.001 29.002 HEAD {PM 9625 Channel C option test} 29.003 DISP UUT Setup: 29.003 DISP Connect UUT Channel C input to 6060 RF output. 29.004 MESS 29.005 MESS Edit 6060 Level Up/Down to Find the stable trigger 29.005 MESS point. 29.006 6060 20mV +20U 150MH -D 30.001 6060 10mV +50% 300MH -D 31.001 6060 10mV +50% 2100MH -D 32.001 HEAD PM 9697 External Referance Multiplier 32.002 OPBR Does the UUT have a Ext Ref Multiplier (?? Option) 32.003 JMPF 33.001 32.004 HEAD {PM 9697 External Referance Multiplier} 32.005 MESS UUT Setup: 32.005 MESS Select PRESET, ENTER, EXT REF 32.005 MESS 32.005 MESS Connect 6060 RF output to the UUT Channel A input 32.005 MESS and the EXT REF input (Rear). Terminate with a 32.005 MESS 50 ohm load. 32.005 MESS 32.006 6060 10MH 0.707V -D S 32.007 EVAL Does the {UUT Display 1.00000000e6 +/-5 Hz}? 33.001 END # IEEE Section 33.002 HEAD {PM6685 Functional Test - IEEE} 33.003 HEAD {Referance Oscillator Check} 33.004 DISP 33.004 DISP Connect UUT Ext Ref input (Rear) to House standard. 33.004 DISP Connect UUT Ch A input to 6060 RF output. 33.004 DISP 33.005 IEEE [TERM LF]*RST 33.006 OPBR Does the UUT have a PM 9678B (02 option) ? 33.007 JMPF 34.002 33.008 6060 10MH 0.707V -D S 33.009 IEEE :READ?[I][D500];:READ?[I] 33.010 MEMC 10000000H 15.0U 34.001 JMP 37.001 34.002 OPBR Does the UUT have a PM 9690 (04 option) ? 34.003 JMPF 35.002 34.004 6060 10MH 0.707V -D S 34.005 IEEE :READ?[I][D500];:READ?[I] 34.006 MEMC 10000000H 2.00U 35.001 JMP 37.001 35.002 OPBR Does the UUT have a PM 9691 (05 option) ? 35.003 JMPF 36.002 35.004 6060 10MH 0.707V -D S 35.005 IEEE :READ?[I][D500];:READ?[I] 35.006 MEMC 10000000H 1.00U 36.001 JMP 37.001 36.002 6060 10MH 0.707V -D S 36.003 IEEE :READ?[I][D500];:READ?[I] 36.004 MEMC 10000000H 150U 37.001 HEAD {Sensitivity and Frequency Range Tests} #IMP 50 ohms, Maximum Sensitivity, Ext Ref on. 37.002 IEEE :INP:IMP 50;:ROSC:SOUR EXT;:INIT:CONT[10] 37.003 IEEE [GTL] 37.004 DISP 37.005 MESS Edit 6060 Level Up/Down to Find the stable trigger 37.005 MESS point. 37.006 6060 10.0mV +50% 1MH -D 38.001 6060 10.0mV +50% 25MH -D 39.001 6060 10.0mV +50% 50MH -D 40.001 6060 15.0mV +50% 100MH -D 41.001 6060 20.0mV +50% 150MH -D 42.001 6060 30.0mV +50% 200MH -D 43.001 6060 50.0mV +50% 300MH -D 44.001 HEAD {Measuring Functions (PRESET)} 44.002 IEEE [@6060]FR10MZ,AP10DB 44.003 DISP 44.003 DISP Connect 6060 to Ch A & Ch E (Ext Arm on rear) 44.003 DISP using a BNC Tee. 44.004 IEEE [TERM LF]*RST;:READ?[I][D500];:READ?[I] 44.005 MEMC 10000000H 50U 45.001 HEAD {Measuring Functions (IMP A 50 ohms)} 45.002 IEEE [@6060]FR10MZ,AP10DB 45.003 IEEE :INP:IMP 50;:READ?[I][D500];:READ?[I] 45.004 MEMC 10000000H 50U 46.001 HEAD {Measuring Functions (Non AUTO)} 46.002 IEEE [@6060]FR10MZ,AP10DB 46.003 IEEE :INP:LEV:AUTO OFF 46.004 MEMC 10000000h 50U 47.001 HEAD {Measuring Functions (PERiod A)} 47.002 IEEE [@6060]FR10MZ,AP10DB 47.003 IEEE :FUNC 'PER 1';:READ?[I][D500];:READ?[I] 47.004 MATH MEM=MEM*1E9 47.005 MEMC 100.000nS 10% 48.001 HEAD {Measuring Functions (RATIO A/E)} 48.002 IEEE [@6060]FR10MZ,AP10DB 48.003 IEEE :FUNC 'FREQ:RAT 1,5';:READ?[I][D500];:READ?[I] 48.004 MEMC 1.000 10% 49.001 HEAD {Measuring Functions (PulseWIDTH A)} 49.002 IEEE [@6060]FR10MZ,AP10DB 49.003 IEEE :FUNC 'PWID 1';:READ?[I][D500];:READ?[I] 49.004 MATH mem=mem*1E9 49.005 MEMC 50.000nS 10% 50.001 HEAD {Measuring Functions (TOT A MAN, DISPLAY HOLD)} 50.002 IEEE [@6060]FR10MZ,AP10DB 50.003 IEEE :FUNC 'TOT 1' 50.004 IEEE [GTL] 50.005 MESS UUT Setup: Select DISPLAY HOLD 50.006 EVAL Does {UUT Start counting (TOT A MAN, DISPLAY HOLD)} ? 51.001 MESS UUT Setup: Select DISPLAY HOLD 51.002 EVAL Does {UUT Stop counting (DISPLAY HOLD)} ? 52.001 HEAD {Measuring Functions (DUTY FACTOR)} 52.002 IEEE [@6060]FR10MZ,AP10DB 52.003 IEEE :FUNC 'PDUT 1';:READ?[I][D500];:READ?[I] 52.004 MEMC .5000 10% 53.001 HEAD {Measuring Functions (AUTO)} 53.002 IEEE [@6060]FR10MZ,AP10DB 53.003 IEEE :INP:LEV:AUTO ON;:READ?[I][D500];:READ?[I] 53.004 MEMC .5000 10% 54.001 HEAD Channel C Options 54.002 OPBR Does the UUT have a Channel C option? 54.003 JMPF 64.001 54.004 OPBR Does the UUT have a 1.3 Ghz channel C option (PM 9621)? 54.005 JMPT 54.010 54.006 OPBR Does the UUT have a 2.7 Ghz channel C option (PM 9624)? 54.007 JMPT 58.002 54.008 OPBR Does the UUT have a 4.5 Ghz channel C option (PM 9625)? 54.009 JMPT 61.002 54.010 HEAD {PM 9621 Channel C option test} 54.011 IEEE [TERM LF]:FUNC 'FREQ 3';:ROSC:SOUR EXT 54.012 IEEE [GTL] 54.013 DISP Connect UUT Channel C input to 6060 RF output. 54.013 DISP 54.014 DISP Edit 6060 Level Up/Down to Find the stable trigger 54.014 DISP point. 54.015 6060 10mV +50% 70MH -D 55.001 6060 15mV +50% 900MH -D 56.001 6060 15mV +50% 1100MH -D 57.001 6060 40mV +50% 1300MH -D 58.001 JMP 64.001 58.002 HEAD {PM 9624 Channel C option test} 58.003 IEEE [TERM LF]:FUNC 'FREQ 3';:ROSC:SOUR EXT 58.004 IEEE [GTL] 58.005 DISP UUT Setup: 58.005 DISP Connect UUT Channel C input to 6060 RF output. 58.005 DISP 58.006 MESS Edit 6060 Level Up/Down to Find the stable trigger 58.006 MESS point. 58.007 6060 20mV +50% 100MH -D 59.001 6060 10mV +50% 300MH -D 60.001 6060 10mV +50% 2100MH -D 61.001 JMP 64.001 61.002 HEAD {PM 9625 Channel C option test} 61.003 IEEE [TERM LF]:FUNC 'FREQ 3';:ROSC:SOUR EXT 61.004 IEEE [GTL] 61.005 DISP Connect UUT Channel C input to 6060 RF output. 61.005 DISP 61.006 DISP Edit 6060 Level Up/Down to Find the stable trigger 61.006 DISP point. 61.007 6060 20mV +50% 150MH -D 62.001 6060 10mV +50% 300MH -D 63.001 6060 10mV +50% 2100MH -D 64.001 HEAD PM 9697 External Referance Multiplier 64.002 OPBR Does the UUT have a Ext Ref Multiplier (?? Option) 64.003 JMPF 33.001 64.004 HEAD {PM 9697 External Referance Multiplier} 64.005 6060 10MH 0.707V -D S 64.006 MESS UUT Setup: 64.006 MESS Select PRESET, ENTER, EXT REF 64.006 MESS 64.006 MESS Connect 6060 RF output to the UUT Channel A input 64.006 MESS and the EXT REF input (Rear). Terminate with a 64.006 MESS 50 ohm load. 64.006 MESS 64.007 EVAL Does the {UUT Display 1.00000000e6 +/-5 Hz}? 65.001 END